sertraline side effects read here- Ranges ±0.1 to ±30 inH2O (0.25 to 70 m bar)
- Accuracy: ±0.5% FS BSL (±0.25% FS optional)
- Long term stability: > ±0.5% FS / Year
- Calibration certificate option
- Wet/wet differentialpressure
- User accessible damping
- Unidirectional or bidirectional operation
- Current or voltage output
Pressure Measurement
Operating Ranges | 0.1, 0.2, 0.25, 0.5, 0.75, 1, 2, 2.5, 3, 4, 5, 6, 8, 10, 16, 20 inH2O and 1 psi unidirectional and bidirectional (other pressure units available, see ordering information). |
---|---|
Overpressure |
|
Maximum Static (Line) Pressure | 30 psi for all ranges |
Electrical Supply Voltage |
Power Supply (Vs) (at the terminals)
Power Supply Intrinsically Safe Units |
Output Signal Option |
Unidirectional
Current
Bidirectional
Current Note: voltage outputs are three wire, current outputs are two wire. |
Zero Offset Adjustment | ±5% FS |
Span Adjustment | ±5% FS |
Response Time | User-adjustable 10 ms to 2 s (factory default setting 10 ms) |
Load Impedance | Voltage output 5 kO minimum Current output RL (kO) < (Vsupply - Vs) 20 |
Performance
Combined Effects of Non-linearity, Hysteresis and Repeatability | ±0.5% FS Best Straight Line (BSL ) ±0.25% FS BSL (optional) |
---|---|
Static Pressure Effects | Effects may be zeroed out at operating line pressure |
Stability | At standard conditions will not change by more than 1% FS averaged over one year |
Supply Sensitivity | <0.05% FS per volt |
Compensated Temperature Range | 32 to 122° F (0 to 50° C) |
Temperature Effects |
Over the range 32 to 122 °F (0 to 50° C)
|
Environmental
Protection | Type 3 IP 64 (Except with C2 screw terminals) |
---|---|
Temperature Range | Operating 32 to 122 °F (0 to 50 °C) |
Humidity | 0 to 100% RH, non condensing. |
Mounting Position Effect |
No effect on span, possible zero shift correctable by potentiometer adjustment. Note: Factory calibrated with diaphragm in the vertical plane. |
Physical
Housing | Aluminum |
---|---|
Wetted Parts | Gases and liquids compatible with aluminum, stainless steel 316L, beryllium copper, brass, Inconel X750, adhesives and plastics. |
Electrical Connection |
|
Pressure Connection | 10-32 UNF female to 0.125 in (3.175 mm) barbed fitting |
Weight: | 1 to 2 lb (0.45 to 0.91 kg) approx depending upon configuration |
Certifications | CE marked ATEX Intrinsically Safe (optional, on current output model only) LCIE05ATEX6081X |